Method Development Report

MS1 Resolution1.0
MS2 Resolution0.75
Calibration MethodC:\MassLynx\IntelliStart\Results\Quantitative\Calibration_20120326_1.cal
Tune MethodC:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\SYSTEM2.ipr
Cone Voltage Range2 - 100
Collision Energy Range2 - 80
Lowest Fragment Mass40.00
Excluded Losses18.00, 44.00

Date: Generated on Thu 19 Apr 2012 at 16:11

Results

IntelliStart generated the following experiments:

MRM Experiment C:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\AICAR_neg.exp

IntelliStart found the following compounds:

Compound Formula/Mass Parent m/z Cone Voltage Daughters Collision Energy Ion Mode
AICAR 338
1
2
337.03
337.03
44
44
78.94
96.90
30
30
ES-
ES-

Compound

AICAR

(338)

Transition 1: ES-, m/z 337.03 -> 78.94

Cone Voltage Optimization (m/z 337.03)
Optimized MS Spectrum (at Cone Voltage 44V)
Collision Energy Optimization (m/z 337.03 -> 78.94)
Optimized Daughter Spectrum (at Collision Energy 30eV)

Transition 2: ES-, m/z 337.03 -> 96.90

Cone Voltage Optimization (m/z 337.03)
Optimized MS Spectrum (at Cone Voltage 44V)
Collision Energy Optimization (m/z 337.03 -> 96.90)
Optimized Daughter Spectrum (at Collision Energy 30eV)