Method Development Report

MS1 Resolution1.0
MS2 Resolution0.75
Calibration MethodC:\MassLynx\IntelliStart\Results\Quantitative\Calibration_20120323_2.cal
Tune MethodC:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\SYSTEM2.ipr
Cone Voltage Range2 - 100
Collision Energy Range2 - 80
Lowest Fragment Mass40.00
Excluded Losses18.00, 44.00

Date: Generated on Fri 23 Mar 2012 at 11:47

Results

IntelliStart generated the following experiments:

MRM Experiment C:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\ATP_pos.exp

IntelliStart found the following compounds:

Compound Formula/Mass Parent m/z Cone Voltage Daughters Collision Energy Ion Mode
ATP 507
1
2
508.03
508.03
36
36
136.12
410.09
46
24
ES+
ES+

Compound

ATP

(507)

Transition 1: ES+, m/z 508.03 -> 136.12

Cone Voltage Optimization (m/z 508.03)
Optimized MS Spectrum (at Cone Voltage 36V)
Collision Energy Optimization (m/z 508.03 -> 136.12)
Optimized Daughter Spectrum (at Collision Energy 46eV)

Transition 2: ES+, m/z 508.03 -> 410.09

Cone Voltage Optimization (m/z 508.03)
Optimized MS Spectrum (at Cone Voltage 36V)
Collision Energy Optimization (m/z 508.03 -> 410.09)
Optimized Daughter Spectrum (at Collision Energy 24eV)