Method Development Report

MS1 Resolution1.0
MS2 Resolution0.75
Calibration MethodC:\MassLynx\IntelliStart\Results\Quantitative\Calibration_20120326_1.cal
Tune MethodC:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\SYSTEM2.ipr
Cone Voltage Range2 - 100
Collision Energy Range2 - 80
Lowest Fragment Mass40.00
Excluded Losses18.00, 44.00

Date: Generated on Fri 11 May 2012 at 14:39

Results

IntelliStart generated the following experiments:

MRM Experiment C:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\Isocitrate_pos.exp

IntelliStart found the following compounds:

Compound Formula/Mass Parent m/z Cone Voltage Daughters Collision Energy Ion Mode
Isocitrate 192 No transitions found.

Compound

Isocitrate

(192)

No parent found. Potential causes can include:

(1) Inconsistent ion beam.

(2) Poor ion intensity.

Ensure concentration of analyte is sufficient and the ion beam is stable.

Following chromatogram shows acquired data for the requested parent mass:

Cone Voltage Optimization (m/z 193.22)
Optimized MS Spectrum (at Cone Voltage 46V)

Potential issues include:

(1) No parent mass or poor intensity observed.

(2) Interference on parent.