XEVO-TQ-S Resolution Optimisation

Reference File(s) UsedC:\MassLynx\Ref\Resolution pos ion.ref
Tune File UsedC:\MassLynx\IntelliStart\Results\Quantitative\Resolution_20120326_2.ipr
Date of Optimisation26-03-2012
Time of Optimisation04 : 46 : 40 PM
Tolerance+/- 0.150 Da

Date: Generated on Mon 26 Mar 2012 at 16:46

MS1 Optimisation

Scan Type : MS1 Scan

Target Peak Width : 1.0000 Da

Positive Ion Mode

Mass Intensity Peak Width Result
102.1300 2.08e+007 0.9611 Pass
772.4610 4.96e+007 0.9855 Pass
1372.0379 1.48e+007 0.9368 Pass
1971.6149 7.74e+006 0.9857 Pass

Final Tune Page Settings

LM Resolution1 : 2.55

HM Resolution1 : 14.64

Ion Energy1 : -0.1

MS2 Optimisation

Scan Type : MS2 Scan

Target Peak Width : 0.7500 Da

Positive Ion Mode

Mass Intensity Peak Width Result
102.1300 1.26e+007 0.7386 Pass
772.4610 3.74e+007 0.8137 Pass
1372.0379 8.86e+006 0.7567 Pass
1971.6149 3.91e+006 0.7457 Pass

Final Tune Page Settings

LM Resolution2 : 2.80

HM Resolution2 : 14.75

Ion Energy2 : 0.5

Resolution Optimisation Check : PASS