Method Development Report

MS1 Resolution1.0
MS2 Resolution0.75
Calibration MethodC:\MassLynx\IntelliStart\Results\Quantitative\Calibration_20120326_1.cal
Tune MethodC:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\SYSTEM2.ipr
Cone Voltage Range2 - 100
Collision Energy Range2 - 80
Lowest Fragment Mass40.00
Excluded Losses18.00, 44.00

Date: Generated on Thu 29 Mar 2012 at 16:00

Results

IntelliStart generated the following experiments:

MRM Experiment C:\MassLynx\Xevo TQ-S - Acquity UPLC Installation Project.pro\ACQUDB\URACIL_pos.exp

IntelliStart found the following compounds:

Compound Formula/Mass Parent m/z Cone Voltage Daughters Collision Energy Ion Mode
URACIL 112
1
2
112.97
112.97
96
96
84.06
56.05
14
14
ES+
ES+

Compound

URACIL

(112)

Transition 1: ES+, m/z 112.97 -> 84.06

Cone Voltage Optimization (m/z 112.97)
Optimized MS Spectrum (at Cone Voltage 96V)
Collision Energy Optimization (m/z 112.97 -> 84.06)
Optimized Daughter Spectrum (at Collision Energy 14eV)

Transition 2: ES+, m/z 112.97 -> 56.05

Cone Voltage Optimization (m/z 112.97)
Optimized MS Spectrum (at Cone Voltage 96V)
Collision Energy Optimization (m/z 112.97 -> 56.05)
Optimized Daughter Spectrum (at Collision Energy 14eV)